0 Rates

Nanowires, nanofibers, nanorods

Nanowires have properties that can’t usually be found in bulk materials due to their energy levels, and that are therefore promising for the preparation of sensors or flexible transparent electrodes, but are also used for numerous other electronic applications. Nanowires, nanofibers, and nanorods also come into play in biomedical applications, where they are used to modify surfaces in order to provide better interaction with biological cells/tissues. In this section, you will find an overview of specific research examples with different measurement techniques, namely microwave synthesis and X-ray diffraction from Anton Paar. Besides the in-situ monitoring of the formation of CuO nanowires with an X-ray diffractometer, you can also read about the formation of nanowires with microwave synthesis.  

Synthesis of copper nanowires for electronic devices

Introduction

Metal nanowires are promising materials for the preparation of sensors or flexible transparent electrodes. Especially nanocopper has attracted considerable interest due to its high electrical conductivity. Various methods in aqueous media have already been investigated to efficiently generate the desired copper nanoparticles. Microwave irradiation facilitates the reductive hydrothermal approach to form the desired nanowires by utilizing inexpensive, readily available biomass as a reducing agent.

Experimental

In a beaker copper (II), chloride and octadecylamine were dissolved in 30 mL water at 65 °C and sonicated for 1 h. Glucose was dissolved in 30 mL water and added to the precursor solution under stirring. An aliquot (max. 20 mL) of the mixture was transferred to a G30 vial and subjected to microwave irradiation. After cooling, the mixture was centrifuged, and the precipitate was thoroughly washed and dispersed in chloroform for storage until purification and further use. Purification of the particles was achieved by simply extracting the chloroform dispersion with water, and centrifugation of the organic phase.

Results and discussion

Different temperature/time conditions (2-4 h at 80 °C to 120 °C) were employed to determine the highest aspect ratio of the resulting nanoparticles. Glucose acts as reducing agent whereas the alkylamine capping agent was essential for the formation of nanowires. However, nanowire formation was not observed in a reaction time under 4 h, and below 120 °C. Conventional heating requires 24 h refluxing to achieve comparable results. The resulting nanowires were used to fabricate transparent conductive electrode films, which remained stable under ambient conditions and acid treatment.

Additional information

Instruments:

Source:

A. S. Hashimi et al., Curr. Appl. Phys. 2020, 20, 205–211

Application Database Entry:

Synthesis of doped hydroxyapatite nanorods with biological activities

Introduction

Porous hydroxyapatite is of considerable interest in nanomaterial research since its porosity and surface area can be easily tuned according to the desired applications in catalysis, gas adsorption, and other fields. Doping with appropriate transition metals like silver or copper leads to remarkable antibacterial activity which makes the material also suitable for the fabrication of medical devices like bone or tooth implants. Microwave irradiation provides a facile and rapid two-step process to generate valuable doped hydroxyapatite nanorods.

Experimental

Calcium nitrate was dissolved together with cetyltrimethylammonium bromide (CTAB) as a surfactant in 30 mL water. Diammonium hydrogen phosphate was separately dissolved in 30 mL water and the pH was adjusted with hydroxylamine to 10. This solution was added dropwise under stirring to the calcium solution. An aliquot (20 mL) was filled in a G30 vial and was subjected to microwave irradiation. Afterwards, the precipitate formed was centrifuged, washed with water and calcined at 550 °C.

For the next step 2 g freshly prepared hydroxyapatite nanorods were dispersed in 50 mL water. 5 mol% silver nitrate was added and the mixture was stirred at room temperature for 5 min. An aliquot (20 mL) was filled in a G30 vial and was subjected to microwave irradiation. The solid material was centrifuged, washed with water and dried at 50 °C.

Results and discussion

Various organic surfactants have been used to study the influence on the morphology of the resulting nanoparticles. The crystal structure of the products was analyzed by powder X-ray diffraction. Solely cationic CTAB as a surfactant already showed good results, but the highest surface area and pore diameter were obtained when a mixture of cationic and anionic surfactants was employed. The doped nanosized hydroxyapatite nanorods generated were evaluated for adsorption properties towards dye and metal ions and antibacterial activity against E. coli. For the latter, a 5 % silver doping showed the highest effectiveness after 24 h of incubation.

Additional information

Instruments:

Source:

M. Sharma et al., J. Nanosci. Nanotechnol. 2018, 18, 623-633

Application Database Entry:

Formation of GeSn nanowires for optoelectronic devices

Introduction

Semiconductor nanowires have attracted interest in recent years as potential material for optoelectronic devices. Incorporation of dopants to the base semiconductor allows for easy tuning of the electronic properties. Especially nanoalloys of tin and germanium in various constitutions show promising effects for LEDs or biological sensors. Microwave heating drastically shortens the reaction time for the formation of desired nanowires and thus allows for simply varying the tin content in the prepared nanomaterial.

Experimental

In a G10 vial bis[bis(trimethylsilyl)]tin and 4 equivalents bis[bis(trimethylsilyl)]germanium were suspended under inert atmosphere in dodecylamine. The mixture was heated at 100 °C under stirring in a glove box overnight. After the pretreatment, the vial was sealed and subjected to microwave irradiation.

After cooling, the mixture was diluted with toluene and centrifuged. The decanted nanowires were redispersed and centrifuged (2x in toluene, 3x in ethanol, 3x in toluene) and finally stored in toluene.

Before further use, the nanowires were annealed under helium atmosphere at 250 °C for 60 min.

Results and discussion

Although rather time-consuming, the pretreatment of the reaction mixture at 100 °C was essential, as otherwise no nanowires would have formed. The nanostructures obtained were characterized by scanning electron microscopy, transmission electron microscopy, energy-dispersive X-ray spectroscopy, and X-ray diffraction (see Fig. 1). Following the described procedure, the resulting nanowires had a high tin content of approx. 20 %. The material was thoroughly evaluated for ohmic behavior by deposition on silicon substrates and revealed strong semiconductor conditions with increasing resistivity with decreasing temperatures.

Figure 1: SEM image (a) and STEM-EDX analysis of GeSn nanowires

Additional information

Instruments:

Source:

M. Sistani et al., Nanoscale 2018, 10, 19443-19449

Application Database Entry:

Synthesis of functionalized nanorods for electronic devices

Introduction

Creating monolayers can be effectively achieved by self-assembly of appropriate substrates, such as functionalized nanorods. Beneficial are self-assembly techniques resulting in thin monolayers, which can be applied in electronic and photonic devices. Here, a microwave-mediated method to efficiently generate metal oxide nanorods, wrapped by reduced graphene oxide, for further self-assembly studies, is presented.

Experimental

In a beaker, graphene oxide sheets were dispersed in 20 mL water and sonicated for 1 h. Anatase powder was added and the mixture was placed on a shaker overnight. The dispersion was then added to 30 mL of a 10M NaOH solution. The resulting mixture was evenly distributed to four 100 mL PTFE liners. The reaction vessels were sealed and placed in the rotor accordingly and subjected to microwave irradiation.

Results and discussion

The parallel microwave approach was chosen to generate sufficient nanomaterial for further investigations in a single experiment. The nanorods were suspended in deionized water (10 mg/mL) and subsequently several coating methods, such as dip-coating, drop-casting or blade coating, were investigated. With the different methods to deposit the as-prepared graphene coated sodium titanate nanorods on various substrates, the accessible thickness of the resulting layers was investigated. The deposited films were characterized by applying the 2D Fourier transformation to scanning electron microscopy images.

Additional information

Instruments:

  • Multiwave PRO (predecessor of Multiwave 5000) with Rotor 8NXF100

Source:

M. H. Modarres et al., Adv. Mater. Interf. 2019, 6, 1900219

Application Database Entry:

Synthesis of core/shell silver-tin oxide nanowires for fabrics

Introduction

Silver nanowires with their remarkable infrared reflectance are of considerable interest as thermal management products for textiles. Current research is focusing on improvements of the nanomaterial stability. Passivation of the nanowires with a nanosized tin oxide layer has potential to enhance the environmental stability while retaining its IR reflectance. Microwave irradiation facilitates a straightforward hydrothermal process to generate core/shell silver-tin oxide nanowires.

Experimental

In a beaker, 3 mL of an ethanolic dispersion of silver nanowires was diluted with water to 100 mL, and sodium citrate (1 wt% in water) was added under stirring. Sodium stannate trihydrate (0.25 wt% in water) was added and the mixture was evenly distributed to eight 100 mL PTFE liners. The reaction vessels were placed in the rotor, which was closed with the lid and subjected to microwave heating at 250 W. The temperature was held at 100 °C for 15 min before further heating to a target temperature of 150 °C.  

Results and discussion

The final nanocomposite was fully characterized by scanning transmission electron microscopy, transmission electron microscopy, electron energy loss spectroscopy, X-ray photoelectron spectroscopy, and UV/Vis spectroscopy. The tin oxide coating improved the air stability of the nanowires, since no degradation could be observed even after 4 months (see Figure 1 below). The nanocomposite was used to produce fabrics for para-aramid textiles to save energy and maintain warmth in the human body.

Figure 1: STEM images of SnO2 decorated nanowires: as synthesized (a), after 3 weeks (b), after 4 months (c)

Additional information

Instruments:

  • Synthos 3000 (predecessor of Multiwave 5000) with Rotor 16MF100

Source:

A. Baranowska-Korczyc et al., RSC Adv. 2021, 11, 4174–4185

Application Database Entry:

Monitoring the formation of CuO nanowires using in-situ high temperature X-ray diffraction

Introduction

Nanostructured semiconductors have a wide variety of potential applications ranging from photovoltaic devices (e.g. solar cells) to gas / liquid sensors. Cuprous oxide (Cu2O) and cupric oxide (CuO) are two naturally occurring oxide forms of copper which can grow as nanoscale crystals. Both display semiconducting properties and their band gaps (2.0 and 1.2 eV, respectively) mean that they can be applied in various fields. Of particular interest are nanowires of the two oxide forms.

Several methods are possible for the growth of CuO nanowires, but the simplest relies on only heating a pure Cu foil in air to a sufficiently high temperature for a certain length of time. The temperature at which nanowire growth occurs is in the region from 300 – 800 °C. 

Experimental

A 0.015 mm thick Cu foil was used as a substrate from which to grow CuO nanowires. In-situ X-ray diffraction (XRD) experiments were performed in air on a laboratory powder X-ray diffractometer in transmission geometry using CuKα radiation. The sample conditions were controlled with the TTK 600 Low-Temperature Chamber with a transmission sample holder from Anton Paar.

The Cu foil was heated from room temperature to 600 °C in 10 °C steps and held at the highest temperature for 4 hours. Measurements were performed for each temperature step and then every 15 minutes at 600 °C. An additional measurement where a foil was rapidly heated to 600 °C at a rate of 100 °C/min and a measurement performed every 2 minutes at 600 °C for 4 hours, with further measurements every 15 minutes for two hours at 600 °C, was also performed.

Results and discussion

The formation of two forms of Cu oxide could be monitored in-situ using the TTK 600 Low-Temperature Chamber from Anton Paar. It was found that on heating a pure Cu foil to 600 °C, Cu2O forms first at ~ 360 °C before the nanowire form of CuO is produced at temperatures above 500 °C. When held at 600 °C, the majority of CuO is formed from Cu2O within the first two hours, though the fraction of CuO present continues to rise over a longer period of time.

This study gives an indication of the temperatures and time scales required to optimize production of nanostructured CuO. In particular, shorter time scales than those often reported in the literature (~ 2 hours, as opposed to > 4 hours) may be sufficient to produce CuO nanowires in suitable quantities.