43 Rates

Density Measurement Applications

The density of liquids is an important physical parameter that plays a critical role not only in consumer protection, trade, safety and healthcare, taxation, and environmental protection but also in research and development.

Accurate density measurement is key for characterizing oils and fuels, for example, as well as determining alcohol concentration in alcoholic beverages for fiscal purposes or sugar content in non-alcoholic beverages. 

It’s also essential for quantifying large amounts of liquid goods when converting a mass flow into a volume flow or vice versa.[1]

Density-based concentration determination

One of the most important applications of density measurement is determining the concentration of binary solutions (i.e., mixtures of two components, such as alcohol-water solutions or sugar/water solutions). 

The density of a solution depends on the concentration of the dissolved substance (“solute”); if the interrelationship is known, the density can be measured and then converted into the respective concentration. A typical application is the determination of sugar content in the soft drink industry. By determining the density of a soft drink, which is considered a sugar-water solution, you can also determine its sugar content.

Concentration determination is also possible when a mixture contains several components and only one of them varies while all other ingredients are constant. 

Modern digital density meters today often include numerous chemical concentration tables for the most important industrial compounds, such as those below.

Acids

NameConc. Range (% w/w)Accuracy Approx. (%) Δ
acetic acid (CH3COOH)0–600.09
citric acid anhydr. (C6H8O7)0–300.02
citric acid hydr. (C6H8O7·H2O)0–30.60.02
EDTA anhydr. (C10H16N2O8)0–60.02
EDTA hydr. (C10H16N2O8·2H2O)0–6.640.02
formic acid (CH2O2)0–700.04
hydrochloric acid (HCl) *0–400.02
lactic acid (C3H6O3)0–800.04
malic acid (C4H6O5)0–70.10.02
nitric acid (HNO3)0–1000.07
oxalic acid (C2H2O4)0–80.02
phosphoric acid (H3PO4)0–1000.06
sulfuric acid (H2SO4)0–930.05
tartaric acid (C4H6O6)0–580.02
trichloroacetic acid (C2HCl3O2)0–500.02

* Not available for Combined Density and Sound Velocity Meters 
Δ The specified accuracy is based on the underlying literature tables

Concentration of different aqueous acids in mol/L according to different tables, based on true density at 20 °C:

NameConc. Range (mol/L)
hydrochloric acid (HCl)0-13.1
nitric acid (HNO3)0-24
sodium hydroxide (NaOH)0-19.07
phosphoric acid (H3PO4)0-19.07
sulfuric acid (H2SO4)0-17.6

Bases

Concentration of different aqueous bases in percentage by weight (% w/w) according to different tables, based on true density at 20 °C:

NameConc. Range (% w/w)Accuracy Approx. (%) Δ
ammonia (NH3)0–300.03
ammonium hydroxide (NH4OH)0–61.740.06
potassium hydroxide (KOH)0–500.01
sodium hydroxide (NaOH)0–500.04

Δ The specified accuracy is based on the underlying literature tables

Alcohols

Concentration of different aqueous alcohols in percentage by weight (% w/w) according to different tables, based on true density at 20 °C:

NameConc. Range (% w/w)Accuracy Approx. (%) Δ
ethylene glycol (C2H6O2)0–600.08
glycerol (C3H8O3)0–1000.04
methanol (CH3OH)0–1000.05
1-propanol (C3H8O)0–1000.05
2-propanol (C3H8O)0–1000.05
propylene glycol (C3H8O2)0–600.14

Δ The specified accuracy is based on the underlying literature tables

Ethanol

Ethanol concentrations in percentage by volume (% v/v) or percentage by weight (% w/w) according to ethanol concentrations tables issued by different organizations/authors:

NameDescription
ethanol Canadian excise (% v/v)Special ethanol table converting the output of a special adjustment into an ethanol value
ethanol AOAC 60 °F (% v/v)Percentage by volume at 15.56 °C (60 °F), American Organization of Analytical Chemists (AOAC) tables, based on true density at 20 °C. The measuring temperature has to be 20 °C (68 °F).
ethanol AOAC 60 °F (% v/v) (not visc.-corr.)Percentage by volume at 15.56 °C (60 °F), American Organization of Analytical Chemists (AOAC) tables, based on true density at 20 °C. The measuring temperature has to be 20 °C (68 °F), without viscosity correction.
ethanol HM C&E (% v/v)HM C&E table at 20 °C
ethanol HM C&E (% w/w)HM C&E table at 20 °C
ethanol IUPAC (% v/v)Tables of the International Union of Pure and Applied Chemistry (IUPAC), based on true density at 20 °C. The measuring temperature has to be 20 °C (68 °F).
ethanol IUPAC (% w/w)Tables of the International Union of Pure and Applied Chemistry (IUPAC), based on true density at 20 °C. The measuring temperature has to be 20 °C (68 °F).
ethanol Kaempf (% v/v)Percentage by volume according to W. Kaempf, based on true density at 20 °C
ethanol Kaempf (% w/w)Percentage by weight according to W. Kaempf, based on true density at 20 °C
ethanol OIML (% v/v)According to the tables of the International Organization of Legal Metrology (OIML), temperature according to ITS 68, based on true density at 20 °C
ethanol OIML (% w/w)According to the tables of the International Organization of Legal Metrology (OIML), temperature according to ITS 68, based on true density at 20 °C
SHUSEI-DO (% v/v)Percentage by volume at 15 °C (59 °F), based on true density at 15 °C (59 °F). The measuring temperature has to be 15 °C (59 °F).
ethanol OIML-ITS-90 (% v/v)According to the tables of the International Organization of Legal Metrology (OIML), temperature according to ITS 90, based on true density at 20 °C
ethanol OIML-ITS-90 (% w/w)According to the tables of the International Organization of Legal Metrology (OIML), temperature according to ITS 90, based on true density at 20 °C
ethanol Proof 60 °F (EtOH °Proof)Proof degrees at 15.56 °C (60 °F), based on true density at 20 °C
ethanol Proof 60 °F (EtOH °Proof) (not visc.-corr.)Proof degrees at 15.56 °C (60 °F), based on true density at 20 °C, without viscosity correction

Inorganic salts

Concentration of different inorganic salts in percentage by weight (% w/w) according to different tables, based on true density at 20 °C:

NameConc. Range (% w/w)Accuracy Approx. (%) Δ
ammonium chloride (NH4Cl)0–240.03
ammonium sulfate ((NH4)2SO4)0–400.02
barium chloride anhydr. (BaCl2)0–260.01
barium chloride hydr. (BaCl2·2H2O)0–300.01
cadmium chloride (CdCl2)0–560.01
cadmium sulfate (CdSO4)0–400.01
calcium chloride anhydr. (CaCl2)0–400.01
calcium chloride hydr. (CaCl2·H2O)0–52.990.01
cesium chloride (CsCl)0–640.01
cobaltous chloride anhydr. (CoCl2)0–200.01
cobaltous chloride hydr. (CoCl2·6H2O)0–36.60.02
cupric sulfate anhydr. (CuSO4)0–180.01
cupric sulfate hydr. (CuSO4·5H2O)0–28.160.01
dipotassium phosphate anhydr. (K2HPO4)0–80.01
dipotassium phosphate hydr. (K2HPO4·3H2O)0–10.480.02
disodium phosphate anhydr. (Na2HPO4)0–5.50.01
disodium phosphate hydr. (Na2HPO4·7H2O)0–10.390.02
ferric chloride anhydr. (FeCl3)0–400.01
ferric chloride hydr. (FeCl3·6H2O)0–66.660.02
lanthanum nitrate anhydr. (La(NO3)3)0–440.01
lanthanum nitrate hydr. (La(NO3)3·6H2O)0–58.640.01
lithium chloride (LiCl)0–300.02
magnesium chloride anhydr. (MgCl2)0–300.01
magnesium chloride hydr. (MgCl2·6H2O)0–640.02
magnesium sulfate anhydr. (MgSO4)0–260.01
magnesium sulfate hydr. (MgSO4·7H2O)0–53.240.02
manganous sulfate anhydr. (MnSO4)0–200.01
manganous sulfate hydr. (MnSO4·H2O)0–22.390.01
monopotassium phosphate (KH2PO4)0–100.01
monosodium phosphate anhydr. (NaH2PO4)0–400.01
monosodium phosphate hydr. (NaH2PO4·H2O)0–460.01
nickel sulfate anhydr. (NiSO4)0–60.01
nickel sulfate hydr. (NiSO4·6H2O)0–10.290.02
potassium bicarbonate (KHCO3)0–240.01
potassium biphthalate (C8H5KO4)0–80.02
potassium bromide (KBr)0–400.01
potassium carbonate anhydr. (K2CO3)0–500.01
potassium carbonate hydr. (K2CO3·1½H2O)0–59.780.01
potassium chloride (KCl)0–240.01
potassium chromate (K2CrO4)0–300.01
potassium dichromate (K2Cr2O7)0–100.01
potassium ferricyanide anhydr. (K3Fe(CN)6)0–300.02
potassium ferrocyanide anhydr. (K4Fe(CN)6)0–200.01
potassium ferrocyanide hydr. (K4Fe(CN)6·3H2O)0–22.930.02
potassium iodide (KI)0–400.01
potassium nitrate (KNO3)0–240.01
potassium oxalate anhydr. (K2C2O4)0–140.01
potassium oxalate hydr. (K2C2O4·H2O)0–15.520.01
potassium permanganate (KMnO4)0–60.01
potassium sulfate (K2SO4)0–100.01
potassium thiocyanate (KSCN)0–640.02
silver nitrate (AgNO3)0–400.01
sodium acetate (CH3COONa)0–300.02
sodium bicarbonate (NaHCO3)0–60.01
sodium bromide (NaBr)0–400.01
sodium carbonate anhydr. (Na2CO3)0–150.01
sodium carbonate hydr. (Na2CO3·10H2O)0–40.490.03
sodium chloride (NaCl)0–260.01
sodium citrate anhydr. (Na3C6H5O7)0–360.01
sodium citrate hydr. (Na3C6H5O7·2H2O)0–41.030.01
sodium dichromate anhydr. (Na2Cr2O7)0–600.01
sodium dichromate hydr. (Na2Cr2O7·2H2O)0–68.250.01
sodium ferrocyanide anhydr. (Na4Fe(CN)6)0–150.01
sodium ferrocyanide hydr. (Na4Fe(CN)6·10H2O)0–23.890.02
sodium molybdate anhydr. (Na2MoO4)0–90.01
sodium molybdate hydr. (Na2MoO4·2H2O)0–10.580.01
sodium nitrate (NaNO3)0–400.01
sodium sulfate anhydr. (Na2SO4)0–220.01
sodium sulfate hydr. (Na2SO4·10H2O)0–49.90.02
sodium tartrate anhydr. (C4H4Na2O6)0–280.01
sodium tartrate hydr. (C4H4Na2O6·2H2O)0–33.20.02
sodium thiocyanate (NaCNS)0–560.02
sodium thiosulfate anhydr. (Na2S2O3)0–400.01
sodium thiosulfate hydr. (Na2S2O3·5H2O)0–62.790.02
sodium tungstate anhydr. (Na2WO4)0–90.01
sodium tungstate hydr. (Na2WO4·2H2O)0–10.10.01
strontium chloride anhydr. (SrCl2)0–360.01
strontium chloride hydr. (SrCl2·6H2O)0–60.550.01
trisodium phosphate anhydr. (Na3PO4)0–80.01
trisodium phosphate hydr. (Na3PO4·12H2O)0–18.550.02
zinc sulfate anhydr. (ZnSO4)0–160.01
zinc sulfate hydr. (ZnSO4·7H2O)0–28.50.02

Δ The specified accuracy is based on the underlying literature tables

Others

Concentration of various aqueous solutions in percentage by weight (% w/w) according to different tables, based on true density at 20 °C:

NameConc. Range (% w/w)Accuracy Approx. (%) Δ
acetone (C3H6O)0–100.08
creatinine (C4H7N3O)0–80.04
deuterium oxide (D2O)0–1000.09
1,4-dioxane (C4H8O2)0–600.16
hydrogen peroxide (H2O2)0–1000.02
procaine hydrochloride (C13H21N2O2·HCl)0–600.05
sodium diatrizoate (C11H8I3N2NaO4)0–400.01
tetracaine hydrochloride (C15H24N2O2·HCl)0–120.08
THAM (C4H11NO3)0–400.04
urea (NH2CONH2)0–460.03

Δ The specified accuracy is based on the underlying literature tables

Extract / sugars

Concentration of different aqueous saccharides and sugar alcohols in percentage by weight (% w/w) according to different tables, based on true density at 20 °C:

NameConc. Range (% w/w)Accuracy Approx. (%) Δ
dextrane (C6H10O5)0–100.03
dextrose anhydr. (C6H12O6)0–600.02
dextrose hydr. (C6H12O6·H2O)0–660.02
d-fructose (C6H12O6)0–700.02
inulin ((C6H10O5)x)0–100.03
lactose anhydr. (C12H22O11)0–180.02
lactose hydr. (C12H22O11·H2O)0–18.950.02
maltose anhydr. (C12H22O11)0–600.02
maltose hydr. (C12H22O11·H2O)0–63.160.02
d-mannitol (C6H14O6)0–150.03
sucrose (C12H22O11)0–1000.02

Δ The specified accuracy is based on the underlying literature tables

Material characterization

In industries like the petrochemical industry the density of a product is a very important characterization number or quality parameter. Standards such as those defined by ASTM define a certain density for every product (e.g. fuels, lubricating oils, or crude oils). Density measurement ensures these materials can be characterized and the quality of each product defined. 

Academic research

Density is also commonly used as a standard parameter in academic research to investigate the special physical behavior of a product. In petrochemistry, for example, scientists measure the density of gasoline at different pressures and temperatures in the laboratory to find out how it will react once injected into an engine.

Quality control based on density

Another important application of density measurement is quality control – not only in the soft drink or petrochemical industry but also in the production of alcoholic beverages (beer, spirits, wine), fuels (including biodiesel and bioethanol), cosmetics, personal care products, pharmaceuticals, and many more.

Filling volume control

Whereas some products are quantified by weight, others are filled, paid, or taxed based on their volume. A well-proven procedure is to measure the weight of a fluid product with a balance, and then calculate its filling volume using the sample’s density. This procedure is usually applied where direct volume measurement is not possible (e.g., in the beverage, petroleum, or chemical industries).

The filling volume of a certain sample weight can be calculated based on the sample’s apparent density:

$$ V = { W \over \rho _{app}}$$

Equation 1: The filling volume V is defined as weight in air W divided by apparent density ρₐₚₚ of the sample

Conclusion

Countless analytical methods are used in today’s laboratories, whether in research facilities or industrial plants. Density measurement is employed not only to specify and describe pure substances, but also to determine concentrations of binary mixtures. It provides important information about the composition of mixtures, which is useful during quality control or when determining filling volume during the bottling of a product. 

References

  1. Fehlauer, H., and H. Wolf. 2006. "Density Reference Liquids Certified by the Physikalisch-Technische Bundesanstalt." Measurement Science and Technology 17: 2588–2592.