Cartechnic CT12 plus / pure Cartechnic CT12 plus / diluted VW G13 / pure VW G13 / diluted
Description
Pure Substance
Automotive antifreeze, also called coolant, is the liquid which is used in the radiator and cooling system of an internal combustion engine for vehicles.
The main ingredient in most coolants is ethylene glycol or propylene glycol. Automotive antifreeze keeps the cooling water from freezing in winter and from boiling in summer. A further task is to lubricate the cooling system, such as the water pump. Most antifreezes contain corrosion inhibitors and other additives, like foam depressants.
Diluted Substance
Besides its use in radiator systems, automotive antifreeze mixed with water is also a heat transfer liquid for cooling thermostats or circulators used in labs.
Common mixtures are 50:50 or 40:60 of antifreeze : water. Other mixtures are 30:70 or 20:80 of antifreeze : water (% by volume). Mixtures with lower antifreeze content have also lower viscosity, which leads to less flow resistance and less load of the thermostat's pump.
Such liquids can be used for flow temperatures down to -25°C, depending on the application the thermostat is used for. Below this temperature the flow resistance due to increasing viscosity (and possibly long hoses with quick connectors) is too high for proper cooling performance.
Viscosity Tables – Measurement data
Table CT12+ Pure
Temp. [°C]
Dyn. Viscosity [mPa.s]
Kin. Viscosity [mm²/s]
Density [g/cm³]
20
31.72
28.12
1.1282
15
40.86
36.11
1.1316
10
53.59
47.22
1.135
5
71.72
63.01
1.1383
0
98.15
85.98
1.1416
-5
137.63
120.22
1.1448
-10
198.25
172.68
1.1481
-15
294.07
255.4
1.1514
-20
450.54
390.18
1.1547
-25
715.28
617.58
1.1582
-30
1181.23
1016.9
1.1616
-35
2036.19
1747.8
1.165
-40
3662.42
3135.1
1.1682
CT12+ Automotive Antifreeze pure - kinematic viscosity and density over temperature
Table VW G13 Pure
Temp. [°C]
dyn. viscosity [mPa.s]
kin. viscosity [mm²/s]
Density [g/cm³]
20
45.37
39.72
1.1421
15
59.80
52.21
1.1455
10
80.03
69.66
1.1488
5
109.42
94.98
1.1521
0
153.14
132.55
1.1553
-5
220.00
189.89
1.1586
-10
325.09
279.80
1.1619
-15
495.44
425.19
1.1652
-20
780.88
668.19
1.1686
-25
1276.40
1089.20
1.1720
-30
2166.70
1843.70
1.1752
-35
3841.20
3259.50
1.1784
-40
7138.30
6040.80
1.1817
-45
13914.00
11743.00
1.1848
VW G13 Automotive Antifreeze pure - kinematic viscosity and density over temperature
Table VW G13 | Diluted 60 : 40 Antifreeze : Water (% by vol)
[exact content of antifreeze, considering volume contraction: 63.15 % by vol]
Temp. [°C]
dyn. viscosity [mPa.s]
kin. viscosity [mm²/s]
Density [g/cm³]
20
6.34
5.8
1.0935
15
7.75
7.06
1.0964
10
9.53
8.67
1.0992
5
11.91
10.81
1.1019
0
15.14
13.71
1.1046
-5
19.64
17.73
1.1073
-10
26.04
23.46
1.1099
-15
35.43
31.85
1.1125
-20
49.65
44.53
1.115
-25
71.79
64.24
1.1174
-30
107.48
95.97
1.1199
-35
167.29
149.07
1.1222
-40
271.31
241.24
1.1247
-45
459.01
407.24
1.1271
-50
813.83
720.44
1.1296
VW G13 diluted 60:40 with water (% by volume) - kinematic viscosity and density over temperature
Table CT12+ | Diluted 50 : 50 Antifreeze : Water (% by vol)
[exact content of antifreeze, considering volume contraction: 51.8 % by vol]
Temp. [°C]
Dyn. Viscosity [mPa.s]
Kin. Viscosity [mm²/s]
Density [g/cm³]
20
3.98
3.72
1.0695
15
4.72
4.4
1.0721
10
5.69
5.29
1.0744
5
6.96
6.46
1.0777
0
8.66
8.02
1.0800
-5
10.98
10.15
1.0822
-10
14.23
13.13
1.0842
-15
18.91
17.4
1.0863
-20
25.83
23.74
1.0883
-25
36.34
33.33
1.0903
-30
52.96
48.49
1.0921
-35
80.24
73.35
1.0939
-40
126.95
115.86
1.0957
CT12+ diluted 50:50 with water (% by volume) - kinematic viscosity and density over temperature
Table VW G13 | Diluted 50 : 50 Antifreeze : Water (% by vol)
[exact content of antifreeze, considering volume contraction: 51.48 % by vol]
Temp. [°C]
dyn. viscosity [mPa.s]
kin. viscosity [mm²/s]
Density [g/cm³]
20
4.29
3.98
1.0771
15
5.14
4.76
1.0797
10
6.22
5.75
1.0824
5
7.64
7.04
1.0849
0
9.53
8.77
1.0873
-5
12.14
11.14
1.0897
-10
15.80
14.47
1.0919
-15
21.08
19.27
1.0941
-20
28.90
26.37
1.0962
-25
40.91
37.25
1.0982
-30
59.89
54.44
1.1001
-35
91.03
82.60
1.1020
VW G13 Automotive Antifreeze diluted 50:50 with water (% by volume) - kinematic viscosity and density over temperature
Table CT12+ Diluted | 40 : 60 Antifreeze : Water (% by vol)
[exact content of antifreeze, considering volume contraction: 45.41 % by vol]
Temp. [°C]
Dyn. Viscosity [mPa.s]
Kin. Viscosity [mm²/s]
Density [g/cm³]
20
3.32
3.13
1.0612
15
3.95
3.71
1.0635
10
4.71
4.42
1.0659
5
5.72
5.35
1.0681
0
7.05
6.59
1.0702
-5
8.88
8.28
1.0722
-10
11.42
10.63
1.0741
-15
15.06
14.00
1.0759
-20
20.44
18.96
1.0776
-25
28.51
26.42
1.0793
CT12+ diluted 40:60 with water (% by volume) - kinematic viscosity and density over temperature
Table VW G13 Diluted | 40 : 60 Antifreeze : Water (% by vol)
[exact content of antifreeze, considering volume contraction: 39.07 % by vol]
Temp. [°C]
dyn. viscosity [mPa.s]
kin. viscosity [mm²/s]
Density [g/cm³]
20
3.01
2.84
1.0591
15
3.54
3.34
1.0614
10
4.22
3.96
1.0636
5
5.09
4.78
1.0656
0
6.26
5.86
1.0676
-5
7.85
7.34
1.0694
-10
10.04
9.38
1.0710
-15
13.18
12.29
1.0726
-20
17.76
16.53
1.0740
VW G13 diluted 40:60 with water (% by volume) - kinematic viscosity and density over temperature
Table CT12+ Diluted | 30 : 70 Antifreeze : Water (% by vol)
[exact content of antifreeze, considering volume contraction: 34.42 % by vol]
Temp. [°C]
Dyn. Viscosity [mPa.s]
Kin. Viscosity [mm²/s]
Density [g/cm³]
20
2.49
2.38
1.0466
15
2.92
2.79
1.0484
10
3.45
3.29
1.0504
5
4.15
3.94
1.0521
0
5.07
4.81
1.0537
-5
6.31
5.97
1.0568
-10
8.05
7.61
1.0574
-15
10.51
9.93
1.0584
-20
14.13
13.34
1.059
CT12+ diluted 30:70 with water (% by volume) - kinematic viscosity and density over temperature
Table VW G13 Diluted | 30 : 70 Antifreeze : Water (% by vol)
[exact content of antifreeze, considering volume contraction: 29.64 % by vol]
Temp. [°C]
dyn. viscosity [mPa.s]
kin. viscosity [mm²/s]
Density [g/cm³]
20
2.34
2.24
1.0450
15
2.70
2.58
1.0470
10
3.18
3.04
1.0488
5
3.81
3.63
1.0505
0
4.63
4.41
1.0520
-5
5.75
5.46
1.0533
-10
7.28
6.90
1.0544
-15
9.45
8.95
1.0553
VW G13 diluted 30:70 with water (% by volume) - kinematic viscosity and density over temperature
Table CT12+ Diluted | 20 : 80 Antifreeze : Water (% by vol)
[exact content of antifreeze, considering volume contraction: 23.21 % by vol]
Temp. [°C]
Dyn. Viscosity [mPa.s]
Kin. Viscosity [mm²/s]
Density [g/cm³]
20
1.86
1.80
1.0313
15
2.16
2.09
1.0327
10
2.53
2.45
1.0342
5
3.01
2.91
1.0355
0
3.65
3.52
1.0366
-5
4.50
4.34
1.0374
-10
5.67
5.46
1.0381
CT12+ diluted 20:80 with water (% by volume) - kinematic viscosity and density over temperature
Table VW G13 Diluted | 20 : 80 Antifreeze : Water (% by vol)
[exact content of antifreeze, considering volume contraction: 19.10 % by vol]
Temp. [°C]
dyn. viscosity [mPa.s]
kin. viscosity [mm²/s]
Density [g/cm³]
20
1.76
1.71
1.0288
15
2.01
1.95
1.0305
10
2.34
2.27
1.0319
5
2.76
2.67
1.0331
0
3.32
3.21
1.034
-5
4.07
3.93
1.0347
-10
5.09
4.92
1.0351
VW G13 diluted 20:80 with water (% by volume) - kinematic viscosity and density over temperature
Table VW G13 Diluted | 10 : 90 Antifreeze : Water (% by vol)
[exact content of antifreeze, considering volume contraction: 12.74 % by vol]
Temp. [°C]
dyn. viscosity [mPa.s]
kin. viscosity [mm²/s]
Density [g/cm³]
20
1.53
1.50
1.0188
15
1.72
1.69
1.0202
10
1.97
1.93
1.0213
5
2.29
2.24
1.0222
0
2.72
2.66
1.0228
-5
3.31
3.23
1.0230
VW G13 diluted 10:90 with water (% by volume) - kinematic viscosity and density over temperature
Volume Contraction for Diluted Substance Antifreeze : Water
Due to volume contraction, the nominal relation values for the mixtures in % by volume are not exact in the above tables. Volume contraction in chemistry means that if mixing two different liquids the total volume is less than the sum of the single components' volumes. For example, if adding 100 mL of water to 100 mL of ethanol, the resulting volume is less than 200 mL. How strong this effect is depends on the components, but it is considerable if mixing water and alcohol.
The exact content of antifreeze in the tested mixtures in % by volume is given in brackets [ ].
See also the charts that show viscosity and density over exact concentration for several temperatures and both substances, CT12+ and VW G13.
CT12+ and VW G13 diluted with water - kinematic viscosity over concentration (% by volume) at several temperatures
CT12+ and VW G13 diluted with water - density over concentration (% by volume) at several temperatures
Reference
Measured with SVM™.
Metainformation
Reference
Measurement with SVM™
Kin. vis.
yes
Dyn. vis.
yes
Density
yes
Multiple temperatures
yes
Related
Glycol, antifreeze, coolant, heat transfer liquid
쿠키 설정
당사는 웹 사이트에서 쿠키를 사용합니다. 일부 쿠키(예: 장바구니)는 필수이며, 일부 쿠키는 분석, 외부 미디어 및 마케팅 서비스를 통해 장기적으로 더 효율적이고 편리한 온라인 콘텐츠를 구축하도록 도움을 줍니다. 저희는 또한 귀하께서 동의하신 경우 Google 이나 Facebook 등 미국 제공자의 제삼자 쿠키를 이용합니다(Art. 49(1)(a) GDPR). 귀하의 데이터는 미국으로 전달되며 미국은 GDPR에 상응하는 데이터 보호가 적용되지 않습니다. 이 경우 미국 당국에서 감독 목적에 따라 귀하의 데이터에 접근할 가능성이 있으며 귀하는 효과적인 법적 구제조치를 이용하지 못할 수 있습니다. 각 버튼을 클릭하여 모든 쿠키를 수락 또는 거부하거나 "쿠키 설정 사용자 지정" 링크를 사용하여 쿠키 설정을 정의할 수 있습니다.