Cartechnic CT12 plus / pure Cartechnic CT12 plus / diluted VW G13 / pure VW G13 / diluted
Description
Pure Substance
Automotive antifreeze, also called coolant, is the liquid which is used in the radiator and cooling system of an internal combustion engine for vehicles.
The main ingredient in most coolants is ethylene glycol or propylene glycol. Automotive antifreeze keeps the cooling water from freezing in winter and from boiling in summer. A further task is to lubricate the cooling system, such as the water pump. Most antifreezes contain corrosion inhibitors and other additives, like foam depressants.
Diluted Substance
Besides its use in radiator systems, automotive antifreeze mixed with water is also a heat transfer liquid for cooling thermostats or circulators used in labs.
Common mixtures are 50:50 or 40:60 of antifreeze : water. Other mixtures are 30:70 or 20:80 of antifreeze : water (% by volume). Mixtures with lower antifreeze content have also lower viscosity, which leads to less flow resistance and less load of the thermostat's pump.
Such liquids can be used for flow temperatures down to -25°C, depending on the application the thermostat is used for. Below this temperature the flow resistance due to increasing viscosity (and possibly long hoses with quick connectors) is too high for proper cooling performance.
Viscosity Tables – Measurement data
Table CT12+ Pure
Temp. [°C]
Dyn. Viscosity [mPa.s]
Kin. Viscosity [mm²/s]
Density [g/cm³]
20
31.72
28.12
1.1282
15
40.86
36.11
1.1316
10
53.59
47.22
1.135
5
71.72
63.01
1.1383
0
98.15
85.98
1.1416
-5
137.63
120.22
1.1448
-10
198.25
172.68
1.1481
-15
294.07
255.4
1.1514
-20
450.54
390.18
1.1547
-25
715.28
617.58
1.1582
-30
1181.23
1016.9
1.1616
-35
2036.19
1747.8
1.165
-40
3662.42
3135.1
1.1682
CT12+ Automotive Antifreeze pure - kinematic viscosity and density over temperature
Table VW G13 Pure
Temp. [°C]
dyn. viscosity [mPa.s]
kin. viscosity [mm²/s]
Density [g/cm³]
20
45.37
39.72
1.1421
15
59.80
52.21
1.1455
10
80.03
69.66
1.1488
5
109.42
94.98
1.1521
0
153.14
132.55
1.1553
-5
220.00
189.89
1.1586
-10
325.09
279.80
1.1619
-15
495.44
425.19
1.1652
-20
780.88
668.19
1.1686
-25
1276.40
1089.20
1.1720
-30
2166.70
1843.70
1.1752
-35
3841.20
3259.50
1.1784
-40
7138.30
6040.80
1.1817
-45
13914.00
11743.00
1.1848
VW G13 Automotive Antifreeze pure - kinematic viscosity and density over temperature
Table VW G13 | Diluted 60 : 40 Antifreeze : Water (% by vol)
[exact content of antifreeze, considering volume contraction: 63.15 % by vol]
Temp. [°C]
dyn. viscosity [mPa.s]
kin. viscosity [mm²/s]
Density [g/cm³]
20
6.34
5.8
1.0935
15
7.75
7.06
1.0964
10
9.53
8.67
1.0992
5
11.91
10.81
1.1019
0
15.14
13.71
1.1046
-5
19.64
17.73
1.1073
-10
26.04
23.46
1.1099
-15
35.43
31.85
1.1125
-20
49.65
44.53
1.115
-25
71.79
64.24
1.1174
-30
107.48
95.97
1.1199
-35
167.29
149.07
1.1222
-40
271.31
241.24
1.1247
-45
459.01
407.24
1.1271
-50
813.83
720.44
1.1296
VW G13 diluted 60:40 with water (% by volume) - kinematic viscosity and density over temperature
Table CT12+ | Diluted 50 : 50 Antifreeze : Water (% by vol)
[exact content of antifreeze, considering volume contraction: 51.8 % by vol]
Temp. [°C]
Dyn. Viscosity [mPa.s]
Kin. Viscosity [mm²/s]
Density [g/cm³]
20
3.98
3.72
1.0695
15
4.72
4.4
1.0721
10
5.69
5.29
1.0744
5
6.96
6.46
1.0777
0
8.66
8.02
1.0800
-5
10.98
10.15
1.0822
-10
14.23
13.13
1.0842
-15
18.91
17.4
1.0863
-20
25.83
23.74
1.0883
-25
36.34
33.33
1.0903
-30
52.96
48.49
1.0921
-35
80.24
73.35
1.0939
-40
126.95
115.86
1.0957
CT12+ diluted 50:50 with water (% by volume) - kinematic viscosity and density over temperature
Table VW G13 | Diluted 50 : 50 Antifreeze : Water (% by vol)
[exact content of antifreeze, considering volume contraction: 51.48 % by vol]
Temp. [°C]
dyn. viscosity [mPa.s]
kin. viscosity [mm²/s]
Density [g/cm³]
20
4.29
3.98
1.0771
15
5.14
4.76
1.0797
10
6.22
5.75
1.0824
5
7.64
7.04
1.0849
0
9.53
8.77
1.0873
-5
12.14
11.14
1.0897
-10
15.80
14.47
1.0919
-15
21.08
19.27
1.0941
-20
28.90
26.37
1.0962
-25
40.91
37.25
1.0982
-30
59.89
54.44
1.1001
-35
91.03
82.60
1.1020
VW G13 Automotive Antifreeze diluted 50:50 with water (% by volume) - kinematic viscosity and density over temperature
Table CT12+ Diluted | 40 : 60 Antifreeze : Water (% by vol)
[exact content of antifreeze, considering volume contraction: 45.41 % by vol]
Temp. [°C]
Dyn. Viscosity [mPa.s]
Kin. Viscosity [mm²/s]
Density [g/cm³]
20
3.32
3.13
1.0612
15
3.95
3.71
1.0635
10
4.71
4.42
1.0659
5
5.72
5.35
1.0681
0
7.05
6.59
1.0702
-5
8.88
8.28
1.0722
-10
11.42
10.63
1.0741
-15
15.06
14.00
1.0759
-20
20.44
18.96
1.0776
-25
28.51
26.42
1.0793
CT12+ diluted 40:60 with water (% by volume) - kinematic viscosity and density over temperature
Table VW G13 Diluted | 40 : 60 Antifreeze : Water (% by vol)
[exact content of antifreeze, considering volume contraction: 39.07 % by vol]
Temp. [°C]
dyn. viscosity [mPa.s]
kin. viscosity [mm²/s]
Density [g/cm³]
20
3.01
2.84
1.0591
15
3.54
3.34
1.0614
10
4.22
3.96
1.0636
5
5.09
4.78
1.0656
0
6.26
5.86
1.0676
-5
7.85
7.34
1.0694
-10
10.04
9.38
1.0710
-15
13.18
12.29
1.0726
-20
17.76
16.53
1.0740
VW G13 diluted 40:60 with water (% by volume) - kinematic viscosity and density over temperature
Table CT12+ Diluted | 30 : 70 Antifreeze : Water (% by vol)
[exact content of antifreeze, considering volume contraction: 34.42 % by vol]
Temp. [°C]
Dyn. Viscosity [mPa.s]
Kin. Viscosity [mm²/s]
Density [g/cm³]
20
2.49
2.38
1.0466
15
2.92
2.79
1.0484
10
3.45
3.29
1.0504
5
4.15
3.94
1.0521
0
5.07
4.81
1.0537
-5
6.31
5.97
1.0568
-10
8.05
7.61
1.0574
-15
10.51
9.93
1.0584
-20
14.13
13.34
1.059
CT12+ diluted 30:70 with water (% by volume) - kinematic viscosity and density over temperature
Table VW G13 Diluted | 30 : 70 Antifreeze : Water (% by vol)
[exact content of antifreeze, considering volume contraction: 29.64 % by vol]
Temp. [°C]
dyn. viscosity [mPa.s]
kin. viscosity [mm²/s]
Density [g/cm³]
20
2.34
2.24
1.0450
15
2.70
2.58
1.0470
10
3.18
3.04
1.0488
5
3.81
3.63
1.0505
0
4.63
4.41
1.0520
-5
5.75
5.46
1.0533
-10
7.28
6.90
1.0544
-15
9.45
8.95
1.0553
VW G13 diluted 30:70 with water (% by volume) - kinematic viscosity and density over temperature
Table CT12+ Diluted | 20 : 80 Antifreeze : Water (% by vol)
[exact content of antifreeze, considering volume contraction: 23.21 % by vol]
Temp. [°C]
Dyn. Viscosity [mPa.s]
Kin. Viscosity [mm²/s]
Density [g/cm³]
20
1.86
1.80
1.0313
15
2.16
2.09
1.0327
10
2.53
2.45
1.0342
5
3.01
2.91
1.0355
0
3.65
3.52
1.0366
-5
4.50
4.34
1.0374
-10
5.67
5.46
1.0381
CT12+ diluted 20:80 with water (% by volume) - kinematic viscosity and density over temperature
Table VW G13 Diluted | 20 : 80 Antifreeze : Water (% by vol)
[exact content of antifreeze, considering volume contraction: 19.10 % by vol]
Temp. [°C]
dyn. viscosity [mPa.s]
kin. viscosity [mm²/s]
Density [g/cm³]
20
1.76
1.71
1.0288
15
2.01
1.95
1.0305
10
2.34
2.27
1.0319
5
2.76
2.67
1.0331
0
3.32
3.21
1.034
-5
4.07
3.93
1.0347
-10
5.09
4.92
1.0351
VW G13 diluted 20:80 with water (% by volume) - kinematic viscosity and density over temperature
Table VW G13 Diluted | 10 : 90 Antifreeze : Water (% by vol)
[exact content of antifreeze, considering volume contraction: 12.74 % by vol]
Temp. [°C]
dyn. viscosity [mPa.s]
kin. viscosity [mm²/s]
Density [g/cm³]
20
1.53
1.50
1.0188
15
1.72
1.69
1.0202
10
1.97
1.93
1.0213
5
2.29
2.24
1.0222
0
2.72
2.66
1.0228
-5
3.31
3.23
1.0230
VW G13 diluted 10:90 with water (% by volume) - kinematic viscosity and density over temperature
Volume Contraction for Diluted Substance Antifreeze : Water
Due to volume contraction, the nominal relation values for the mixtures in % by volume are not exact in the above tables. Volume contraction in chemistry means that if mixing two different liquids the total volume is less than the sum of the single components' volumes. For example, if adding 100 mL of water to 100 mL of ethanol, the resulting volume is less than 200 mL. How strong this effect is depends on the components, but it is considerable if mixing water and alcohol.
The exact content of antifreeze in the tested mixtures in % by volume is given in brackets [ ].
See also the charts that show viscosity and density over exact concentration for several temperatures and both substances, CT12+ and VW G13.
CT12+ and VW G13 diluted with water - kinematic viscosity over concentration (% by volume) at several temperatures
CT12+ and VW G13 diluted with water - density over concentration (% by volume) at several temperatures
Reference
Measured with SVM™.
Metainformation
Reference
Measurement with SVM™
Kin. vis.
yes
Dyn. vis.
yes
Density
yes
Multiple temperatures
yes
Related
Glycol, antifreeze, coolant, heat transfer liquid
Cài đặt cookie
Chúng tôi sử dụng cookie trên trang web của chúng tôi. Một trong số chúng cần thiết (ví dụ như cho giỏ hàng), số khác giúp chúng tôi cải thiện nội dung trực tuyến tốt hơn và dễ dàng hơn cho bạn trong dài hạn thông qua phân tích, truyền thông bên ngoài và dịch vụ tiếp thị. Chúng tôi cũng sử dụng cookie từ bên thứ ba từ các nhà cung cấp tại Mỹ như Google hoặc Facebook nếu được cung cấp sự đồng ý của bạn (Điều 49(1)(a) GDPR). Dữ liệu của bạn có thể được chuyển đến Hoa Kỳ, nơi không có mức độ bảo vệ dữ liệu tương đương với GDPR. Trong trường hợp này, dữ liệu của bạn có thể bị cơ quan chức năng Mỹ truy cập để theo dõi và bạn có thể không thể thực hiện các biện pháp pháp lý hiệu quả. Bạn có thể chấp nhận hoặc từ chối tất cả cookie bằng cách nhấn vào nút tương ứng hoặc định nghĩa cài đặt cookie của bạn bằng cách sử dụng liên kết 'Tùy chỉnh cài đặt cookie của bạn'.
Nếu bạn từ chối tất cả cookie, chỉ có cookie cần thiết về mặt kỹ thuật sẽ được sử dụng. Bạn cũng có thể rút lại sự đồng ý của mình sau này bằng cách truy cập vào Cài đặt Cookie.
Cookie tiếp thị được sử dụng bởi bên thứ ba hoặc nhà xuất bản để hiển thị quảng cáo cá nhân cho bạn. Họ làm điều này bằng cách theo dõi xem khách truy cập trang web nào.
Chúng tôi thu thập và kết hợp dữ liệu về khách truy cập và hành vi của họ trên trang web của chúng tôi. Thông tin này được sử dụng để cải thiện trang web.
Nội dung từ các nền tảng video và các nền tảng truyền thông xã hội bị chặn mặc định. Nếu bạn chấp nhận cookie từ phương tiện truyền thông bên ngoài, truy cập nội dung này không cần phải đồng ý thủ công nữa.
Sau khi được thông báo về các nguy cơ tiềm ẩn cho dữ liệu của tôi (bao gồm việc chuyển giao đến Mỹ và tiềm ẩn việc truy cứu của các cơ quan chính phủ Mỹ) trong Biểu ngữ Cookie và Chính sách Bảo mật, tôi chấp nhận các cookie từ các nhà cung cấp Mỹ.